Plasma Thawer Use (Saraha III)
Posted: Monday, February 26th, 2024 | Updated: Monday, February 26th, 2024
Posted: Monday, February 26th, 2024 | Updated: Monday, February 26th, 2024
Cleaning:
Use alcohol-based disinfectant for regular cleaning. Unscheduled disinfections may use oxygen generating sporicides.
Use:
Turn Power Switch to on position. This is located on the left side of the machine when facing it
The machine will enter Standby Mode and begin to warm to 37 degrees, but the fan won't begin blowing and the agitator won't start
Place blood bag with all labels and packaging removed on the warming plate underneath the infrared sensor to the left
The bag should lay parallel/vertical to you on the left of the warming plate
Close the lid or system flap and press the start button
Within 30 seconds, the machine will begin operate the agitator, the fan will blow, the temperature of the bag will be sensed, and the plasma will begin thawing
Once the infrared sensor senses the blood product is free of ice, the free of ice LED will permanently light up and an audible alarm will sound every 5 minutes
After 90 minutes of alarming with no action, the machine will sound an alarm again and go into standby mode
Remove the blood product and turn Power Switch to the off position
Main parts:
Power switch, Agitator, Warming Plate, Infrared Sensor, The Housing, Free of Ice Indicator, Fast Tempering Key, LED error light, Display Temperature Alarm, Blood Component Temperature Display, Start Key, Stop Key, Heating On Display, LED Power On Indicator
Use alcohol-based disinfectant for regular cleaning. Unscheduled disinfections may use oxygen generating sporicides.
Use:
Turn Power Switch to on position. This is located on the left side of the machine when facing it
The machine will enter Standby Mode and begin to warm to 37 degrees, but the fan won't begin blowing and the agitator won't start
Place blood bag with all labels and packaging removed on the warming plate underneath the infrared sensor to the left
The bag should lay parallel/vertical to you on the left of the warming plate
Close the lid or system flap and press the start button
Within 30 seconds, the machine will begin operate the agitator, the fan will blow, the temperature of the bag will be sensed, and the plasma will begin thawing
Once the infrared sensor senses the blood product is free of ice, the free of ice LED will permanently light up and an audible alarm will sound every 5 minutes
After 90 minutes of alarming with no action, the machine will sound an alarm again and go into standby mode
Remove the blood product and turn Power Switch to the off position
Main parts:
Power switch, Agitator, Warming Plate, Infrared Sensor, The Housing, Free of Ice Indicator, Fast Tempering Key, LED error light, Display Temperature Alarm, Blood Component Temperature Display, Start Key, Stop Key, Heating On Display, LED Power On Indicator